We will provide safe and high performance ultra-high pressure devices that meet your needs.|PRETECH CO., LTD.

Low-Temperature X-ray Fluorescence Holography Device XFME-RR4K

Features

By using a cryogenic refrigerator, it is possible to cool the sample surface of a rotary device for low-temperatures to cryogenic temperatures.

●By operating the low-temperature rotary device in the cryogenic temperature, the sample can be rotated under cryogenic temperature environment.
●The low-temperature rotary base can be connected to the refrigerator unit by various attachments, and the attachment can rotate the sample at each temperature, including the temperature sensor and heater.
Main Specifications
Cryogenic refrigerator
ConfigurationRefrigerator unit, Compressor unit
Combined refrigeration capacityTwo-stages: 0.5 W at 4.2 K (50/60 Hz)
One-stage: 3/4 W at 50 K (50/60 Hz)
Cooling methodWater-cooled: Water volume (3 L/min ~ 3.5 L/min @ 28 ºC)
WeightRefrigerator unit: About 15 kg, Compressor unit: About 70 kg
Rotary device for low-temperature
Operating range360°
Temperature range4.2 K to 300 K
Maximum load1 N
Rotary supportSize: Sample surface; ø30, Height; 15.2 mm
Cryostat
Main raw materialsStainless steel, Aluminum, Copper
AttachmentsFor rotary base mounting Parts for low-temperature (copper)
※ Built-in temperature sensor and heater
Radiation insulationTube material for radiant heat insulation (steel)
Other attachmentsVacuum chamber (stainless steel, Aluminum)
Vacuum exhaust valve, Wiring connector